RECONDITIONED OEC 9900 ELITE C-ARM

  • The Elite & Elite CFD are both more expensive new then their new GE OEC 9900 counter parts, with Flat Detector models starting in the high $100 thousands and refurbished systems are simply not available yet do to them just being released the end of last year (on average its take 3-5 years before you’ll see refurbished systems available).
  • The OEC 9900 Elite is designed to deliver superb clinical images during both simple and challenging imaging cases, with fast and easy positioning. See which configuration – based on II size, type of C and optional features – will best meet your clinical needs. Imaging profiles optimize your image to compensate for bone and tissue.
9900

.GE OEC 9900 C-Arm & GE OEC Elite CFD C-Arm Are The Same Except For The Monitor Cart And Flat Detector. Now let’s dig into those changes deeper. 9900 II vs Elite II vs Elite CFD Flat Detector: When comparing the GE OEC 9900 to the GE OEC Elite, they both come in 9″ and 12″ Image intensifier models. GE OEC 9900 Elite C-Arm With Continual High Level Fluoroscopy. GE is a trusted brand that has saturated the medical equipment market. They produce intuitive, well-working medical equipment. C-arms are large pieces of imaging equipment that take captures of inside the body.

The reconditioned OEC 9900 Elite select configurations are ideal for Office,Outpatient and Ambulatory Center settings.

Superb Imaging
• 1k x 1k x 16 bit resolution imaging
• Advanced image processing accentuates what needs to be seen and attenuates what isn’t needed
• Dynamic Range Management technology processes images such as bone and vasculature separately for optimum definition
• Dose management options for more control: Digital Spot,High Level Fluoroscopy, Standard Fluoroscopy, Low Dose, and Pulse Mode
• Smart Metal for enhanced imaging when metal is present
• AutoTrak sampling window automatically moves with the anatomy to ensure superb images during live fluoro

Workflow Efficiency
• Simple user interface with touchscreen reduces procedure time to improve efficiency
• Flat panel monitors on an articulating arm allow viewing of images from anywhere in the room
• Superb mechanical C-arm design features counterbalancing and friction controls for easier AP and lateral movement, orbital rotation, and head-to-toe angulation

Oec 9900 elite service manualUser

Refurbishing Process:
All OEC-9800’s go through a multistep refurbishing process. Each unit is restored to a like-new status. Pain Management, Orthopedics and Rehabilitation Services are just a few of the applications utilizing the refurbished OEC-9800 c-arm. The rotating x-ray tube allows for greater penetration and resolution for the lumbar area when doing needle localizations or other work involving the thicker anatomical areas of patients. For barium swallows a VCR or Digital Video Recorder can be utilized for better record keeping and post study analysis. The 1K X 1K CCD Technology used in the OEC-9800 image system allows for crisp, clear imaging. Whatever the purpose, a refurbished OEC-9800 will deliver the results you can depend on.

Oec 9900 Elite User Manual

Mechanical repairs/cosmetic maintenance
• All defective parts and covers are repaired or replaced
• C-arm is completely disassembled and repainted
• “C” is removed from main frame assembly
• Horizontal arm removed, inspected and repaired as necessary
• Vertical column removed and bearings removed inspected and replaced as required
• X-ray tube removed and inspected for artifacts and internal integrity
• Image intensifier removed and inspected for any imperfections and grid integrity
• Camera removed
• Wheels removed, cleaned and or replaced from main chassis
• Chassis inspected for structural defects and repaired as necessary
• Vertical lift assembly is tested for drift and adjusted/replaced if necessary
• All bearings are inspected, cleaned and re-lubricated, or replaced as needed
• Control panel display is tested and replaced if necessary
• All wheels are inspected and lubricated and/or replaced as needed, with new cable pushers installed
• All locks and brake assemblies are inspected, cleaned, and repaired or replaced as needed
• C-arm is completely disassembled and repainted
• Steering is tested and adjusted is required
• Foot switch and hand switch provided on each unit
• A complete set of operator and service manuals are provided

Oec 9900 Elite User Manual

Oec 9900 elite user manual instructions

Re-assembly
• All components sent to C-arm reassembly area for rebuilding
• “C”, Image tube and camera, Vertical column, main control and power cable installed
• New batteries are installed
• New CRT monitors installed in monitor cart
• Monitor cart rebuilt
• When C-arm is completely rebuilt mechanically system is sent for calibration and final testing

Testing/calibration
• Image intensifier and X-ray tube are tested for stability and balance
• Image intensifier is tested for resolution and gain to be within OEM specifications
• X-ray tube bearings are tested for noise and coast time
• X-ray tube filaments and stator windings are tested
• X-ray tube radiation output is verified to be within OEM specifications
• X-ray generator high voltage tested
• Maximum dose rate is set in compliance with government regulations
• CCD camera calibrated to OEM specifications
• X-ray beam is aligned for each field size
• C-arm is tested for current leakage requirements
• All monitors are aligned, resized, centered and focused
• All monitors are adjusted for brightness and linearity
• Video system is optimized for gray scale and resolution
• All power supplies are tested and calibrated or replaced as necessary

Ge Oec Elite 9900 Pdf

Additional Information

Additional Information

Specifications

X-ray System

Generator
• 60kHz high frequency
• 15kW power
• Up to 120kVp
• Up to 75mA for radiographic film exposure
• Continuous high level fluoro (HLF) up to 20mA
• Digital spot up to 75mA
• Digital cine pulse
- 15 or 30 pulses per second, 60 Hz
- 12 or 25 pulses per second, 50 Hz
- Up to 150mA
- 10ms pulse width
• Full power from standard wall outlet
• Patented battery buffered design


X-ray Tube
• Rotating anode X-ray tube

• 0.3 and 0.6 nominal focal spots
• Anode heat capacity: 300,000 HU
• Anode cooling rate: 85,000 HU/min.
• Housing heat capacity: 1,600,000 HU
• Housing cooling rate:
- 9” I.I. Super C- 22,500 HU/min
- 12” I.I. - 22,500 HU/min


Digital Image Rotation
• Digitally adjusts image display
• Automatic image update
- Image rotation
- Image reversal (side-to-side)
- Image invert (top-to-bottom)
• Image positioning without additional exposure


PreView Collimator
• On-screen collimator position indication
• PreView iris collimator
• PreView Tungsten rotatable double leaf collimator
• Adjusts collimators without X-ray exposure

Fluoro Mode
• kVp range: 40 -120
• mA range: 0.2 - 10 normal mode 1.0 - 20 HLF (high level fluoro)
• Auto and manual fluoro modes
• AutoTrak ABS varies mA, kVp, camera gain

Pulsed Fluoro Mode
• kVp range: 40 - 120
• mA range: 0.2 - 10
• Pulse rate: 1, 2, 4, 8
• Pulse width: 25 or 50ms
• AutoTrak ABS, mA, kVp, camera gain
• Reduces X-ray dose to patient and operator

High Level Pulsed Fluoro
• kVp range: 40 - 120
• mA range: 1 - 40
• Pulse rate: 1, 2, 4, 8
• Pulse width: 25 or 50ms
• AutoTrak ABS, mA, kVp, and camera gain

Digital Spot Mode
• kVp range: 40 - 120
• mA range: Up to 75
• Automatic exposure termination and automatic image save Radiographic Mode
• mA range: up to 75
• mAs range: up to 300
• Computer controlled exposure time
• Optional film cassette holder
- 10” x 12” (24cm x 30cm) for 9” I.I.
- 14” x 14” (35cm x 35cm) for 12” I.I.

Video Imaging System 9” Image Intensifier
• Tri-mode 9”/6”/4.5” (23cm/15cm/11cm) image intensifier
• Minimum central resolution (at monitor):
- 9” (23cm): 2.2 lp/mm
- 6” (15cm): 3.0 lp/mm
- 4.5” (11cm): 3.5 lp/mm
• DQE: 65% (typical) 12” Image Intensifier
• Tri-mode 12”/9”/6” (31cm/23cm/15cm) image intensifier
• Minimum central resolution (at monitor):
- 12” (31cm): 1.6 lp/mm
- 9” (23cm): 2.2 lp/mm
- 6” (15cm): 2.6 lp/mm
• DQE: 65% (typical)


Precision Imaging with Dynamic Range Management (DRM)
• Preset Imaging Profile: Orthopedic
• Enhances features of interest while attenuating background noise

AutoTrak Automatic Brightness Stabilization (ABS)
• Automatically seeks the subject anatomy anywhere within the imaging field and selects the optimum imaging technique
• Automatically adjusts to anatomical size and location
• Provides uniform image quality throughout entire image
• Simplifies operation

Image Quality Features
• Smart Window
- Dynamically senses the collimator position and automatically adjusts brightness and contrast to produce high image quality
• Smart Metal
- Allows user to adjust automatic brightness and contrast sensitivity levels to metal
- Provides optimum image quality even when metal is introduced to the field
• Tungsten Collimator
- Dense collimator limits X-ray exposure area
- Reduces scatter radiation
- Improves image detail Video Camera
• High resolution 1k x 1k CCD camera
• Full frame capture
• Motorized rotation
• On-screen orientation indicator (real-time feedback without fluoro)
• Left-right image reversal
• Top-bottom image invert Video Display
• Dual 19” (48cm) display anti-glare, LCD flat panel monitors mounted on an articulating arm
- 22” horizontal travel
- 7° up/10° down
- Monitors viewable from all four sides of workstation
- Horizontal and vertical viewing angle 170°

• 1200 CD/M2 maximum brightness
• Touch screen system control
• 1280 x 1024 high resolution monitors
• Integrated color monitor for display of VGA, DVI, DVI-D, S-VHS, and SDISD formats

Software Package Office Pain (OP) and Ambulatory Center (AC)
• 1k x 1k x 16 bit image processing
• Preset Imaging Profile: Orthopedic
• Noise filter with on-screen indicator
• Minimal difference spatiotemporal noise filter (MDST)
• Real-time dynamic range management (DRM)
• Automatic digital brightness and contrast control
• Manual digital brightness and contrast control
• Negate mode
• Save and auto-save feature
• Swap and auto-swap feature
• Patient information
- Examination list
- Customized patient information
• Customize functions
- Workstation set-up
- Mainframe set-up
- Patient information set-up
- Date/time set-up
- DICOM interface set-up
• Last image hold
• 63 image storage
• CD/DVD burner with DICOM viewer for displaying images on PC platforms
- 512 x 512 or 1k x 1k
• Integrated DICOM interface (store, print, worklist, and query/retrieve)
• HIPAA SecureView
- Password protection
- Blank screen function
- Delete all


Ambulatory Center Extended (ACE)
Includes all the AC features and:
• 1000 image storage
• Zoom and roam function
• Image annotation
• Measurement software


Office Pain Extended (OPE)
Includes all the OP features and:
• Real-time digital subtraction (DSA)
• Peak opacification
• 8fps Dynamic digital disk
- Recording/playback rate: 1, 2, 4, 8fps
- Recording time: 5 minutes @ 8fps
- Automatic image playback
- Frame-by-frame review

Additional Features
Hardcopy Options • Integrated film/paper printer
- No film developing required
- 8” x 10” (20.3cm x 25.5cm) laser quality film/paper
- Multi-format, 1, 2, 4 on 1
- Multi-copy capability
• Thermal printers
• Integrated DICOM interface (storage class/print class/query work-list)
User Interface
• Entire system is computer controlled
• Touchscreen control simplifies operation
• Automated system operation requires minimum operator interface

• Multi-functional controls
- Footswitch
- IR remote (optional)
- Hand-held control
• Simplified keyboard with integrated touchpad
• Multi-purpose image directory
- Retrieve and review images
- Compose hardcopy films
- Copy images
• X-ray dose summary

Uninterruptible Power
• Orderly shutdown Electrical
• Input power (60 Hz or 50 Hz) 9900 Elite
- 100V @ 20 A
- 120V @ 12 A
Regulatory Compliance
• U.S. 21 CFR Subchapter J
• NFPA 99
• UL 60601-1 (CSA/NRTL)
• IEC60601-1 (plus relevant Collateral and Particular Standards)

File AttachmentsNo